Assembling and Handling the Entonox Kit
1. Self-Contained Kit
Complete Package:
- The Entonox kit is self-contained in a blue bag.
- When you open the bag, you'll find everything needed for patient use.
- Components include the regulator, mouthpiece, and the gas bottle with Schrader valve.
2. Assembly Process
Step-by-Step Setup:
- Connect the Schrader valve on the bottle to the pipe's Schrader valve and twist until it clicks into place.
- Once assembled, turn on the Entonox set.
- Provide the assembled kit to the patient for use; Entonox is administered as the patient sucks on the tube.
3. Disassembly and Cleaning
Safe Removal:
- After Entonox use, disassemble the kit for cleaning and storage.
- Remove and dispose of the one-time-use mouthpiece.
- Turn off the bottle to disable it.
- Depress the flush valve to release pressure from the pipe and regulator before detaching the pipe.
- Ensure depressurization to prevent accidents when disassembling.
4. Cleaning and Storage
Proper Maintenance:
- Clean the components, including wiping them down with sterile wipes.
- Check the cylinder's remaining contents with the regulator.
- Restow the kit in its bag, reassemble, and prepare it for the next patient.
5. Documentation
Record Usage:
- Document the use of Entonox or Nitronox on the patient report form.
- Provide verbal handover of drug administration to the attending medical crew.